4.3.5 Lubricating the Shutter Module

The shutter module should be lubricated approximately every 2000 strokes at
areas specified as shown in

Figure 4-8

. Failure to lubricate the shutter module may

significantly reduce life of operation or cause assembly to become inoperative.

For lubrication, use molybdenum disulfide (PN 750-191-G1), provided with each
shuttered sensor, or use Fomblin® E25 (perfluorinated polyether), if appropriate for
the process.

Figure 4-8 Lubrication guide

1

2

3

NOTES:

1 - Apply molybdenum disulfide to the coupling stem when
extended of 1.

2 - Apply molybdenum disulfide to the shaft hole of 2.

3 - Apply molybdenum disulfide to the pin and hole of 3.

Advertising