INFICON PGD400 Pirani Gauge Display User Manual

Page 10

Advertising
background image

10

tinb14e1 (2009-07)

Measurement values

Measurement range
(air, O

2

, CO, N

2

)

5×10

-4

… 1000 mbar

(+1.9 … +10.0 VDC)
(

→ [1])

Measurement error

Gain error
Offset error


≤0.02% FSr

≤0.05% FSr

Measurement rate

30 / s

Display rate

50 / s

Filter time constant

150 ms (f

g

= 1 Hz)

Pressure units

mbar, Pa, Torr

Gauge supply

Voltage

+24 VDC

±5%

Current 750

mA

Power consumption

18 W

Fuse protection

900 mA with PTC element, self-
resetting after turning the
PGD400 off or disconnecting the
gauge

Switching function

Number 1
Reaction delay

≤10 ms if switching threshold

close to measurement value (for
larger differences consider filter
time constant).

Adjustment range

1×10

-3

… 500 mbar

Hysteresis

10% of measurement value