2 mains power connector – INFICON PGD400 Pirani Gauge Display User Manual

Page 14

Advertising
background image

14

tinb14e1 (2009-07)

3.2 Mains

Power

Connector

DANGER

DANGER: line voltage
Incorrectly grounded products can be extremely
hazardous in the event of a fault.
Use only a 3-conductor power cable (3×1.5 mm

2

)

with protective ground. The power connector may
only be plugged into a socket with a protective
ground. The protection must not be nullified by an
extension cable without protective ground.

The unit is supplied with a 2 m power cord. If the mains cable is
not compatible with your system, use your own, suitable cable
with protective ground.

The socket must be fuse-
protected with 10 A

max

Disconnecting device

acc. to EN 61010-1

If the unit is installed in a switch cabinet, the mains voltage
should be supplied and turned on via a central power distributor.