4 object structure, 1 connection object – INFICON PSG55x ATM to Medium Vacuum Gauge (DeviceNet) User Manual

Page 11

Advertising
background image

tira58e1-a (2012-03) PxG5xx.cp

11

4 Object

Structure

Class Code 05

h

= 05

d

The connection class allocates and manages the internal resources associated
with both I/O and explicit messaging connections. The specific instance generated
by the connection class is referred to as a connection instance or a connection
object.

The following Instances of the connection object are supported:
• Instance 1: Explicit messaging
• Instance 2: I/O polling
• Instance 4: I/O change of state/cyclic I/O connection

Please refer to the DeviceNet specification for further information.

For the selection of the active input assembly the following vendor specific attribute
can be used.


Vendor specific extension:

Attribute

ID

Access

rule

NV/

V

Name

DeviceNet

Data type

Description of attribute

Semantics

100 Get/Set

NV

Poll

produce

assembly
instance

USINT

Contains the instance num-
ber of the assembly used by
this connection to send data

1, 2 (default), 3, 4, 5, 6,
8, 9, 10, 11, 12, 13, 14

This vendor-specific attribute facilitates the configuration of the data assembly, the
PCG/PSG55x will send to the DeviceNet master as poll response. It offers the
possibility to select a (predefined) data assembly via a configuration tool + EDS file
(

→  10). Attribute 100 allows the user to configure the poll IO data assembly via

EDS even if the poll connection of the PCG/PSG55x is in the established state.
The assembly number is stored non volatile. Modifications of the poll produce
assembly instance will take effect only after a reset of the PCG/PSG55x.
Thus, definition of the PCG/PSG55x data assembly can be done in various ways:
1) Standard method (without using attribute 100):

The DeviceNet master configures the "Produced Connection Path" attribute of
the poll connection (connection instance 2) when establishing communication to
the PCG/PSG55x. This requires a master to support expanded explicit
messaging capabilities.

2) Directly setting attribute 100 e.g. by a configuration tool (e.g. RS networks) +

device reset.