INFICON PSG55x ATM to Medium Vacuum Gauge (DeviceNet) User Manual

Page 35

Advertising
background image

tira58e1-a (2012-03) PxG5xx.cp

35




Attribute

ID

Access rule NV/

V

Name

DeviceNet
data type

Description of attribute

Semantics of values

3 Set

NV

High

trip

point

INT or based
data type
attribute

Defines the value at or
above which a trip point
condition will occur.

→ "Semantics" section below
[default = defined by vendor]

4 Set

NV

High

trip

enable

BOOL

Enables the high trip point
setting.

[default = defined by vendor]

5 Set

NV

Low

trip

point

INT or based
data type
attribute

Defines the value at or be-
low which a trip point con-
dition will occur.

→ "Semantics" section below
[default = defined by vendor]

6 Set

NV

Low

trip

enable

BOOL

Enables the low trip point
setting.

[default = defined by vendor]

7 Get

V

Output
Status

BOOL

State of this object
instance.

0 = trip point condition does not exist
(unasserted)
1 = trip point condition exists
(asserted)

12 Get

NV

Destination

Packed
EPATH

Specifies the path of the
destination attribute whose
value will be set by output.

Path to attribute 3 of Instance 1 or 2
of DOP Object.

13

Get

V Output

BOOL

Output of the object the
value of which is sent to
destination.

= Status of the relay

14 Get

NV

Source

Packed
EPATH

Specifies the path of the
source attribute whose
value is retrieved for Input.

Path to attribute 6 of instance 1 of S-
analog sensor object

15

Get

V Input

INT or speci-
fied by data
type

Input to the object whose
value is retrieved from
source.

Value of attribute 6 of instance 1 of S-
analog sensor object

16

Get

NV Data units

ENGUNITS

Units of input, trip point.

17

Get

NV Data type

USINT

Data type of input.

[default] = INT

18 Set

NV

Trip

Point

High
Hysteresis

Same as
High/Low
Point

Data

Type

Determines the amount by
which the Input must
recover to clear a trip point
high condition

[default = defined by vendor]

19 Set

NV

Trip

Point

Low
Hysteresis

Same as
High/Low
Point

Data

Type

Determines the amount by
which the Input must
recover to clear a trip point
low condition

[default = defined by vendor]

20

Get

V Trip Status USINT

Describes the activator for
the trip point condition.
Activator may be a
High Trip Point condition
and/or a Low Trip Point
condition.

0 = no trip point condition exists
1 = low trip point condition exists
2 = high trip point condition exists
3 = high and low trip condition exist
4-255 = reserved

101 Set

NV

Factor

of

ATM

REAL

Multiplier for ATM control

If with the Attribute 102 the amt mode
is chosen, this value will be multiplied
with pressure value from S-Analog
Sensor Instance 3 (ATM sensor) and
copied into High or Low Trip Point
(

→ Attribute 102).

102 Set

NV

Setpoint
Mode

USINT

Mode of the setpoints

Atm Mode,
0 = standard setpoint
1 = LTP is in ATM Mode
2 = HTP is in ATM Mode
3 = LTP and HTP are in ATM Mode
4 … 7 = the same as 0 … 3, but the
change with push-buttons isn't al-
lowed

4.5.2 Instance Attributes of

Trip Point Objects
Instance 1 & 2

Advertising