tira58e1-a (2012-03) PxG5xx.cp

Service

Code

Service Name

Description of Service

32

h

User Atmosphere
Adjust

Performs an Atmosphere Adjust: Piezo will be
adjusted to the Capacitance Diaphragm
Gauge.

Parameter

Data Type

Semantics of Values

User Atmosphere Adjust
parameter

BOOL

0 = disable adjustment
1 = Execute one User
Atmosphere Adjust

To perform the User Atmosphere Adjust Request the flange input has to be in the
same environment as the electronic part and then start the User Atmosphere
Adjust Service with the value 1.

Following is the instance 1 with the subclass extension of the heat transfer mano-
meter part of the gauge (valid for PSG55x).


Attribute

ID

Access rule NV/

V

Name

DeviceNet
data type

Description of attribute

Semantics of values

3

"Semantics"

NV Data type USINT

Determines the data type
of value and all related
attributes as specified in
this table.

→ "Semantics" section below
[default] = INT

4

"Semantics"

NV Data units UINT

Determines the units con-
text of value and all related
attributes.

[default] = Counts

5 Get

V

Reading
valid

BOOL

Indicates that the value
attribute contains a valid
value.

0 = invalid
1 = valid
(invalid: e.g., sensor is out of its
specified range)

6

Get

V Value

INT or REAL

Analog input value.

The corrected, converted, calibrated
final value of the sensor.
→ "Semantics" section

7

Get

V Status

BYTE

Alarm and warning state of
this object instance.

= 0

10

Get

NV Full scale

INT or REAL

The value of full scale for
the sensor.

[default] = maximum allowable value
for the data type

25

Set

NV Safe state USINT

Specifies the behavior for
the value for states other
than execute.

→ "Semantics" section below
[default] = 0

26

Set

NV Safe value INT or speci-

fied by data
type

The Value to be used for
safe state = safe value

→ "Semantics" section below
[default] = 0

32

Get

NV Overrange INT or REAL

Specifies the highest valid
value.

The value above which attribute
reading valid

is set to invalid.

[default] = maximum allowable value
for the data type

33 Get

NV

Under-
range

INT or REAL

Specifies the lowest valid
value.

The value below which attribute
reading valid

is set to invalid.

[default] = minimum allowable value
for the data type

94 Get

V

Sensor
warning

Structure of
byte

Bit definitions of sensor
warnings.

default = 0
→ "Semantics" section below

95 Get

V

Sensor
alarm

Structure of
byte

Bit definitions of sensor
alarms.

default = 0
→ "Semantics" section below