1 < zero > adjustment – INFICON SKY CDG160D 1 ... 1000 Torr / mbar User Manual

Page 25

Advertising
background image

tina53e1-c (2011-11)

25

4.2.1

<ZERO> Adjustment

The zero can be adjusted via

• the <ZERO> button on the gauge,
• the diagnostic port (→  [4]),
• the digital input "Remote Zero" (briefly apply the

supply voltage to pin 10 (15-pin) or to pin 7 (9-pin)),

• the RS232C interface (→  [4]),
• an INFICON Vacuum Gauge Controller (VGC series).

While the gauge is being heated and/or under atmos-

pheric pressure, the zeroing function is locked in order
for operating errors to be prevented.

Evacuate the gauge to a pressure according to the table

below:

F.S.

Recommended final pressure for

zero adjustment

1100 mbar
1000 Torr/mbar
500 Torr/mbar
200 Torr/mbar
100 Torr/mbar
50 Torr/mbar
20 Torr/mbar
10 Torr/mbar
5

Torr/mbar

2

Torr/mbar

1

Torr/mbar

0.5 Torr/mbar
0.25 Torr/mbar
0.1 Torr/mbar


<5

×10

-2

Torr

<2.5

×10

-2

Torr

<10

-2

Torr

<5

×10

-3

Torr

<2.5

×10

-3

Torr

<10

-3

Torr

<5

×10

-4

Torr

<2.5

×10

-4

Torr

<10

-4

Torr

<5

×10

-5

Torr

<2.5

×10

-5

Torr

<10

-5

Torr

<5

×10

-6

Torr

<6.65

×10

0

Pa

<6.65

×10

0

Pa

<3.33

×10

0

Pa

<1.33

×10

-0

Pa

<6.65

×10

-1

Pa

<3.33

×10

-1

Pa

<1.33

×10

-1

Pa

<6.65

×10

-2

Pa

<3.33

×10

-2

Pa

<1.33

×10

-2

Pa

<6.65

×10

-3

Pa

<3.33

×10

-3

Pa

<1.33

×10

-3

Pa

<6.65

×10

-4

Pa

<5

×10

-2

mbar

<5

×10

-

2 mbar

<2.5

×10

-2

mbar

<10

-2

mbar

<5

×10

-3

mbar

<2.5

×10

-3

mbar

<10

-3

mbar

<5

×10

-4

mbar

<2.5

×10

-4

mbar

<10

-4

mbar

<5

×10

-5

mbar

<2.5

×10

-5

mbar

<10

-5

mbar

<5

×10

-6

mbar

If the final pressure is too high for zero adjustment (>25%
of the F.S.), the zero cannot be reached and the
<STATUS> LED blinks green. If this is the case, activate
the factory setting and adjust the zero again (

→  32).

Advertising
This manual is related to the following products: