Intended

Use

The temperature compensated Capacitance Diaphragm Gauges
of the CDG160D and CDG200D series are intended for absolute
pressure measurement of gases in their respective pressure
ranges (

→  3).

The gauges belong to the SKY

®

Smart Sensors family and can

be operated in connection with an INFICON Vacuum Gauge
Controller (VGC series) or another appropriate controller.

Functional

Principle

A ceramic diaphragm is deflected by pressure. The deflection is
measured capacitively and converted into an analog linear out-
put signal by the digital electronics.
The output signal is independent of the gas type.
Very accurate pressure measurement is achieved by heating the
sensor to a constant temperature of 160°C or 200 °C which re-
sults in a compensation of changes in the ambient conditions
and a reduced deposition of process products and by-products in
process applications. An integrated baffle protects the sensor
against coating.

Trademarks