INFICON SKY CDG160D 1 ... 1000 Torr / mbar User Manual
Page 26
                                    Advertising
                                    
                                        
                                    
                                
                                 
26
tina53e1-c (2011-11)
Operate the gauge for at least 2 hours under constant
ambient conditions (until the signal is stable).
Briefly press the <ZERO> button with a pin (max.
ø1.1 mm). The zero adjustment runs automatically. The 
<STATUS> LED blinks until the adjustment (duration ≤8 s) 
is completed. 
Press the button briefly
max. ø1.1 mm
After zero adjustment, the gauge automatically returns
to the measurement mode.
The <STATUS> LED blinks green if
• the signal output is negative (< -20 mV) when the final pres-
sure has been attained
• the zero adjustment has failed.
4.2.2
<ZERO> Adjustment with Ramp Function
The ramp function allows to adjust the zero at a known reference 
pressure within the measurement range of the gauge. 