INFICON MPG40x ATM to Ultra-High Vacuum Gauge User Manual

Page 11

Advertising
background image

tina48e1-b (2008-10) MPG400-401.om

11




• Allen wrench 1.5 mm
• Open-end wrench 7 mm


3

2

4

7

1

a) Unfasten the hexagon socket set screw (1) on the electronics unit (2).

b) Remove the electronics unit without twisting it.

c) Unfasten the hexagon head screw (3) on the magnet unit (4) and remove the

magnet unit.

The magnetic force and the tendency to tilt make it difficult to

separate the magnet unit and the measuring chamber (7).

d) Make the flange connection between the gauge and the vacuum system.
e) Remount the magnet unit and lock it with the hexagon head screw (3).
f) Carefully mount the electronics unit (2). (Make sure the pin of the Pirani ele-

ment is properly plugged into the corresponding hole of the electronics unit.)

g) Push the electronics unit up to the mechanical stop and lock it with the hexagon

socket set screw (1).



3.1.1 Removing the Magnet

Unit (Only for Gauges
With CF Flanges)

Tools required

Procedure