3 reassembling mpg401 – INFICON MPG40x ATM to Ultra-High Vacuum Gauge User Manual

Page 24

Advertising
background image

24

tina48e1-b (2008-10) MPG400-401.om

(

→ figure 22)

Insert the Pirani element (13) with the screw fitting (13a) and copper seal

(13b) slid over it into the corresponding conic bore hole (7) of the measuring
chamber.

Tighten the screw fitting (13a) with your fingers while slightly pushing the

Pirani element against the mechanical stop. Then tighten the screw fitting
by one turn with the open-end wrench.

Insert a new metal seal (11) incl. the centering ring (12) centered into the

measuring chamber (7).

Carefully insert the anode (10) with the ignition aid (10a) and extension

piece (9a) slid onto it into the measuring chamber.

Carefully place the pressure piece (9) on the measuring chamber.

Insert the four hexagon socket screws (8) incl. lock washers (8a) and

tighten them uniformly until the mechanical stop is reached.

Carefully slide the insulator (9b) onto the pressure piece (9) and lock it with

the hexagon socket set screw (9c).

Position the ignition aid (10a) by pushing the mounting tool over the anode

pin until the mechanical stop is reached.

Blow the particles in the measuring chamber with dry nitrogen (be careful to

hold the measuring chamber with the flange pointing downwards).

Slide the polarity insert (6) into the measuring chamber until the mechanical

stop is reached.

Place the circlip (5) snugly fitting on the polarity insert.

Visually check that the anode pin is centered over the middle

hole of the polarity insert (max. eccentricity = 0.5 mm).

If possible perform a leak test (leak rate <10

-9

mbar l/s). If necessary slightly

retighten the screw fitting (13a).

WARNING

WARNING: electric arcing
Helium may cause electric arcing with detrimental effects on
the electronics of the product.
Before performing any tightness tests put the product out of
operation and remove the electronics unit.

Mount the magnet unit (4) and lock it with the hexagon head screw (3).

Carefully mount the electronics unit (2). (Make sure the pin of the Pirani ele-
ment is properly plugged into the corresponding hole of the electronics unit.)

Push the electronics unit up to the mechanical stop and lock it with the
hexagon socket set screw (1).

Adjust the gauge (

→ 16).

6.3.3 Reassembling MPG401

Advertising