2 cleaning mpg401 – INFICON MPG40x ATM to Ultra-High Vacuum Gauge User Manual

Page 23

Advertising
background image

tina48e1-b (2008-10) MPG400-401.om

23


Using a polishing cloth rub the inside walls of the measuring chamber and

the polarity insert to a bright finish.

The sealing surfaces must only be worked concentrically.

Rinse the measuring chamber and the polarity insert with cleaning alcohol.

Allow both to dry.


Remove the used ignition aid (10a) with pliers.

Using a polishing cloth rub the anode pin to a bright finish.

Do not bend the anode. Do not carry out mechanical work on the

ceramic part.

Rinse the anode with cleaning alcohol.

Allow the anode to dry.

Insert a new ignition aid (10a) into the mounting tool.

Carefully press the anode (clean or new)

centered and parallel to the tool axis into
the ignition aid and insert it to a depth of
approx. 15 mm. The final positioning is
established after the anode is installed.

Ignition aid

Mounting tool

Anode


Slide the screw fitting (13a) and the copper seal (13b) over the Pirani

element (13).

Mount the Pirani element (

→ 24).


6.3.2 Cleaning MPG401

Cleaning the measuring
chamber and the polarity
insert

Cleaning or replacing the
anode

Replacing the Pirani element