2 cleaning mpg400 – INFICON MPG40x ATM to Ultra-High Vacuum Gauge User Manual

Page 19

Advertising
background image

tina48e1-b (2008-10) MPG400-401.om

19

Unfasten the hexagon head screw (3) on the magnet unit (4) and remove

the magnet unit.

The magnetic force and the tendency to tilt make it difficult to

separate the magnet unit and the measuring chamber (7).

Remove the circlip (5) and the polarity insert (6) from the measuring

chamber

Remove the three hexagon socket screws (8) incl. lock washers (8a) on the

back of the measuring chamber.

Carefully remove the following parts in this order (without exerting stress on

the Pirani element (13)): pressure piece (9), complete anode (10), FPM seal
(11) incl. support ring (12), Pirani element (13) incl. FPM seal (13a).

The parts can now be cleaned or replaced individually.


Using a polishing cloth rub the inside walls of the measuring chamber and

the polarity insert to a bright finish.

The sealing surfaces must only be worked concentrically.

Rinse the measuring chamber and the polarity insert with cleaning alcohol.

Allow both to dry.


Remove the used ignition aid (10a) with tweezers.

Using a polishing cloth rub the anode pin to a bright finish.

Do not bend the anode. Do not carry out mechanical work on the

ceramic part.

Rinse the anode with cleaning alcohol.

Allow the anode to dry.

Insert a new ignition aid (10a) into the mounting tool.

Carefully press the anode (clean or new)

centered and parallel to the tool axis into
the ignition aid and insert it to a depth of
approx. 15 mm. The final positioning is
established after the anode is installed.

Ignition aid

Mounting tool

Anode


6.2.2 Cleaning MPG400

Cleaning the measuring
chamber and the polarity
insert

Cleaning or replacing the
anode