5.2.3 Total Measurement Cycle

Figure 3: Measurement Cycle

Ch 1 (yellow) depicts the signal directly at the sensor

Ch 2 (blue) equals the RDY (Ready) signal


Figure 3 shows a measurement cycle. A measurement cycle is defined as the time between two con-
secutive negative edges of the Ready-signal. As can be seen from Figure 3, that time equals 682µs.
The timing source for the measurement cycle is an on chip oscillator. Although it is very stable it
shows a relatively large absolute frequency tolerance of about +/- 15% from chip to chip. Therefore the
cycle time can also vary in the range of about +/-15% from sensor to sensor.


5.2.4 Ready-signal

The Ready-signal is active low and it is set after the first measurement value is available. The Ready-
signal is reset by assigning the SS/-signal active. (SPI Slave Select) that means it is automatically
reset by a consecutive read of the measurement value.