5 appendix, 1 grounding concepts – INFICON Spot CDS550D User Manual
Page 10
Research & Development
SPOT CDS500D & CDS550D OEM Sensor SPI-Interface Specification_V1.1
Page 10 of 14
Christian Berg, T based on Felix Mullis, TL
Created 11/10/2014 2:36:00 PM
5 Appendix
5.1 Grounding Concepts
Situation
•
high impedance (> 100kOhm) between sensor elec-
tronics and
–
housing
•
very sensitive measurement frontend (resolution is
fF)
•
shielding against noise necessary
•
the metallic case of the sensor must be electrically
connected to the power supply reference terminal
(GND) of the driving electronics
Concept #1
•
“indirect” connection via earth lines
•
easy to implement
•
keep impedance of earth connections as low as pos-
sible
Concept #2
•
direct connection which may optionally be connected
to earth
•
more laborious to implement
•
„direct“ connection
better shielding possible