5 appendix, 1 grounding concepts – INFICON Spot CDS550D User Manual

Page 10

Advertising
background image

Research & Development

SPOT CDS500D & CDS550D OEM Sensor SPI-Interface Specification_V1.1

Page 10 of 14

Christian Berg, T based on Felix Mullis, TL

Created 11/10/2014 2:36:00 PM

5 Appendix

5.1 Grounding Concepts

Situation

high impedance (> 100kOhm) between sensor elec-
tronics and

housing

very sensitive measurement frontend (resolution is
fF)

shielding against noise necessary

the metallic case of the sensor must be electrically
connected to the power supply reference terminal
(GND) of the driving electronics


Concept #1

“indirect” connection via earth lines

easy to implement

keep impedance of earth connections as low as pos-
sible


Concept #2

direct connection which may optionally be connected
to earth

more laborious to implement

„direct“ connection

better shielding possible

Advertising