4 history, 3 status and error, 4 power-up reset – INFICON Spot CDS550D User Manual

Page 9: Research & development

Advertising
background image

Research & Development

SPOT CDS500D & CDS550D OEM Sensor SPI-Interface Specification_V1.1

Page 9 of 14

Christian Berg, T based on Felix Mullis, TL

Created 11/10/2014 2:36:00 PM

3.3 Status and Error


The SPI communication string for reading the status value consists of total 4 bytes as follows:

0x48 – 0x00 – 0x00 – 0x00, only the first byte is relevant, the other three must be sent but the content
of them doesn’t matter.


Read Status Op-Code:

Byte_3

Byte_2

Byte_1

Byte_0

0 1 0 0 1 0 0 0

don’t care


Status result:

Byte_3

Byte_2

Byte_1

Byte_0

don’t’ care

24-Bit Status value (Bit 23..0)




Bit

Description

23 SPI communication took place during measurement
13 Pressure error

8

port 3 error

7

port 2 error

6

port 1 error

5

port 0 error

3

Temperature error


All other bits not mentioned here must be ignored.


3.4 Power-up reset

After powering up the device, it must be reset by sending the reset command to the device. The reset
command consists of the single byte 0x88h. It is highly recommended to send the reset command
(0x88h) to the sensor after every power up.

4 History

Ver-
sion

Date

by

Contents

1.0

Jan 17, 2014

Christian

Berg

initial version

1.1

Nov 10, 2014

Felix Mullis

timing specification in section 2.4 and 5.2


Advertising
This manual is related to the following products: