background image

tira92e1-a (2014-07) CDGsci Web Interface.cp

7

• Zoom of high resolution analog output. An arbitrary range can be zoomed.
The lower limit of the selected pressure range is displayed at the analog output with
0.0 V, the upper limit with 10.0 V. The values between these limits are
automatically scaled.

First enter the high mark, then the low mark before setting the zoom to On.
Activating/deactivating the zoom by selecting On or Off. Setting the values takes
approx. 3 s.



This tab offers a wide range of configuration possibilities for the signal flow inside
the Cube.

• Change conversion time capacity/voltage
• Switch temperature compensation on/off
• Switch pressure compensation on/off
• Switch ATM compensation on/off
• Change filter settings
• Change output signal of Temp/ATM
• Change pressure unit


Standard settings for the Options page are:
• C-V Conversion: 62 ms
• T-Comp:

on

• P-Comp:

on

• ATM-Comp:

on

• Adaptive

Filter: Slow

• 2. Stage Filter: None
• Low-Res DAC: SL TempOut
• Unit:

Torr


1.5 Tab <Zoom>

1.6 Tab <Options>