INFICON Cube CDGsci (Web Interface) User Manual

Page 8

Advertising
background image

8

tira92e1-a (2014-07) CDGsci Web Interface.cp

Higher value: Less noise, but slower reaction to
pressure steps.


Activate/deactivate compensation for temperature
fluctuations by selecting On or Off.

Switching this option off will deactivate the
temperature compensation of Cube. The user
has to take care for the correct treatment of the
pressure data.
In an unstable environment, it is recommended
to leave this option at On.


Activate/deactivate compensation for pressure fluctuations
by selecting On or Off.

Switching this option off will deactivate the
linearization of Cube. The output pressure will
resemble a hyperbolic function due to the
nature of the capacitive measurement. The user
has to take care for the correct treatment of the
pressure data.


Activate/deactivate compensation for atmospheric
pressure fluctuations by selecting On or Off.

Switching this option off will deactivate the
atmospheric pressure compensation of Cube.
The user has to take care for the correct treat-
ment of the pressure data.
In an unstable environment, it is recommended
to leave this option at On.


Fast:

The Cube CDGsci responds more quickly to
changes in pressure. As a result, noise will be
larger.

Slow:

The Cube CDGsci responds slower to changes
in pressure. As a result, noise will be smaller.

Dynamic: Depending on the measuring conditions, the

Cube CDGsci selects dynamically <Fast> or
<Slow>.


Conversion time capacity /
voltage

Temperature compensation

Pressure compensation

ATM Compensation

Adaptive Filter