tira92e1-a (2014-07) CDGsci Web Interface.cp

9

In situation of very long static measurements, the data can
be further smoothed by applying a secondary stage filter. If
a Second Stage Filter is applied, the CDC and the
Adaptive Filter settings are adapted accordingly.

Moving Exp AVG: Exponential moving average.
Savitzky-Golay: Symmetrical filter of 4

th

order with

n

L

= n

R

= 5.

Loess: Locally

weighted

scattered data smoothing.

When a Second Stage Filter is applied, the
settings for the C-V Conversion and Adaptive
Filter are fixed to 62ms and Slow respectively.
They cannot be changed as long as a Second
Stage Filter is active.


This setting changes the signal output at <TEMP / ATM>:
Either current ATM pressure or temperature of the gauge
can be output.
For scaling of temperature and pressure see operating
manual (

→  [1]).


Changes pressure unit of the measurement values,
threshold values, etc.



• Continuous logging of data up to 24h, 1s/point
• Download of data possible, even during logging.

<Start> starts the logging, <Stop> ends the logging. The start time/date and the
running number of measurement points are displayed. <Download (.csv)> opens a
download dialogue.


Second Stage Filter

Resolution

Pressure unit

1.7 Tab <Logging>