Example: Parameter setting for line 6

The text in line 6 is selected from the internal text list of the recorder by linking
the internal logic inputs.

Text field (3)

Text field (3) (line 7) can be written to as long as the batch is not completed.
The internal text editor (button

) can be used to input any number

of up to 16 digits. After the batch has been completed, the batch number is
automatically incremented.

Example: Parameter setting for line 7

Parameter for line 6

Parameter setting

Configuration ➔ Report/Batches ➔
Ext.Report/Batches ➔ Batches ➔
Line 6 ➔ Contents right column

Binary-linked text

Configuration ➔ Report/Batches ➔
Ext.Report/Batches ➔ Batches ➔
Line 6 ➔ Binary linking

Logic inp1-2

Configuration ➔ Report/Batches ➔
Ext.Report/Batches ➔ Batches ➔
Line 6 ➔ from text No.

80

Configuration ➔ Report/Batches ➔
Ext.Report/Batches ➔ Batches ➔
Line 6 ➔ Text editable?

Yes

Configuration ➔ Texts ➔ Text 80

normal version

Configuration ➔ Texts ➔ Text 81

heavy version

Configuration ➔ Texts ➔ Text 82

plus version

Configuration ➔ Texts ➔ Text 83

plusC version

Parameter

Configuration … Binary-linked texts

Number of possible texts

Logic inp1-2

4

Logic inp1-3

8

Logic inp1-4

16

Logic inp1-5

32

Logic inp1-6

64

Parameter for line 7

Parameter setting

Configuration ➔ Report/Batches ➔
Ext.Report/Batches ➔ Batches ➔
Line 7 ➔ Contents right column

Batch No.