Operation – Leica Biosystems VT1000 S User Manual

Page 18

Advertising
background image

16

Instructions for Use, V 2.0 RevD – 09/2013

Function:

• When the V-Max button is activated in manual mode (LED on - red light)

and the REV/FORW button is pressed, the knife moves towards the speci-

men at maximum speed.

• When the START button is pressed, the LED in the V-Max button is

extinguished. Sectioning starts at the speed previously selected.

Setting a sectioning window:

5. Operation

2

1

Button with LED

Button with LED

If - accidentally - only one limit stop of the sectioning window is

set, the knife covers the maximum sectioning range!

• Activate the V-Max button. Press REV/FORW toggle switch for fast move-

ment of the blade towards the specimen. Press the button to set the

first limit of the sectioning window.

• Press REV/FORW once again, moving the blade edge past the specimen

block and press once more to set the second sectioning window limit.

• Press START to deactivate V-Max. The knife edge moves back to the

first sectioning window limit and resumes sectioning at the previously

selected speed (10-speed rotary potentiometer).

Function:
• Start single or continuous sectioning stroke – according to whether

SINGLE or CONT mode has previously been selected (see description of

Single/Cont mode for further details).

• Specimen feed (section thickness) takes place prior to each section.

• Retraction (specimen is lowered) takes place when the knife reaches

the rear inversive point.

• In SINGLE mode, the knife stops automatically in the rear end position.

• In CONT mode, START/STOP has to be pressed again to stop the section-

ing movement. The knife stops in the rear end position.

• A sectioning process, once started, will continue.

Fig. 16

Advertising