Leica VT1000 S - Vibrating-blade microtome

Cleaning the knife

When cleaning the knife/blade, always wipe from the knife or blade back towards the cutting

edge, NEVER wipe in the opposite direction - risk of injury!

Clean using an alcohol-based solution or acetone.

7. Cleaning and Maintenance

Before each cleaning, carry out the following preparatory steps:

• Switch off the instrument and disconnect the power plug.

• Remove the blade from the knife holder and insert it in the receptacle at the bottom of the

blade dispenser.

• Remove the knife holder for cleaning.

• Remove the specimen plate from the buffer tray and lay it flat on the stage. Carefully remove

the specimen with a single-edge blade.

• Remove section waste using tweezers or a brush.

• Remove the buffer tray, empty it and rinse it separately with water (

refer also to Chap. 5.6

.)

7.1 Cleaning the instrument

Instrument and outside surfaces

If necessary, the varnished outside surfaces of the control panels can be cleaned with a mild

commercial household cleaner or soap water and then be wiped with a cloth.

The instrument must be completely dry before it can be used again.

Always remove the knife / blade before detaching the knife holder from the instrument.

Always put the knife (blade) back into the knife case or blade dispenser when not in

use!

When using cleaners, comply with the safety instructions from the manufacturer and

the labor-safety regulations at your laboratory.

When cleaning the outer surfaces, do not use xylene or solvents containing acetone or

xylene. The finished surfaces are not resistant to xylene or acetone!

Ensure that liquids do not enter the interior of the instrument during cleaning.