Leica VT1000 S - Vibrating-blade microtome

5.6 Routine daily maintenance and switching off the instrument - VT1000 S

After all daily procedures have been finished, perform the following:

• Switch off the main switch at the back of the instrument.

• Place the magnifier cover on the magnifier.

• Remove the knife holder.

• Take the blade out of the knife holder and dispose it properly and safely.

• Remove the specimen disc and lay it flat on the stage.

• Remove the specimen using a single-edge blade. Then, remove remains

of cyanoacrylate adhesive from the specimen disc.

• Remove and empty out the buffer tray. Dispose of the contents of the

buffer tray properly.

• Drain the cooling bath.

To do so, release the tube from its holder at the rear of the instrument

and dispose of the contents of the ice bath into a suitable vessel. Then

wipe off with a dry cloth.

5. Operation

Caution! The contents of the ice bath can become contaminated if buffer solution is

spilled over it.