4 adjusting the amplitude, Adjusting the amplitude, Operation – Leica Biosystems VT1000 S User Manual

Page 21

Advertising
background image

19

Leica VT1000 S - Vibrating-blade microtome

Adjusting the retraction
• In programming mode, press DISP to display the specimen retraction

menu.

• Display: "LO".
• Set specimen retraction between 1 and 999 µm via the -/+ button; or dis-

able by selecting "0".

• The selected value will be displayed in the FEED window.
• Press CLR to quit the menu function.

5.4 Adjusting the amplitude

• To obtain excellent sectioning results, the amplitude requires adjustment

according to the specimen type being sectioned.

5. Operation

1

Fig. 21

When adjusting the amplitude setting, do not remove the clamping screw, simply

loosen it.

The instrument is shipped with the amplitude set to 0.6 mm.

To this end:
• With a No. 2.5 Allen key loosen the clamping screw (1) and secure the

eccentric on the bottom with your finger.

Selectable amplitude positions are, from left to right:

0.2 mm; 0.4 mm; 0.6 mm; 0.8 mm; 1 mm.

• Slide the amplitude clamping screw to the desired amplitude position

and retighten.

Fig. 22

Advertising