Malfunctions: meaning and troubleshooting – Leica Biosystems VT1000 S User Manual

Page 29

Advertising
background image

27

Leica VT1000 S - Vibrating-blade microtome

6. Malfunctions: Meaning and Troubleshooting

Error messages/symptoms

Sources of error

Troubleshooting

- Audible warning signal.

- Error code E0.8xx is displayed.

- E-EPROM defective.

- Instrument can still be used,

though there will be certain

limitations:

all values will be set to default

values.

New values (sectioning win-

dow, feed, lowering) cannot be

saved.

- Call the Technical Service.

- Audible warning signal.

- Optical signal via red LED.

- The upper limit of the specimen

feed has been reached.

- Leave the upper limit position

(Switch the UP/DOWN button in

DOWN direction).

- Mount a new specimen onto

the specimen holder and start

again.

- Audible warning signal.

- Optical signal via red LED.

- The lower limit of the specimen

level has been reached (height

adjustment of specimen via buf-

fer tray).

- After unlocking the DOWN posi-

tion the buffer tray is automati-

cally raised until the audible and

optical signals turn off.

- Audible warning signal.

- User has tried to select a speci-

men thickness via the "+/-" but-

ton that is below the minimum

value (0 µm) or above the maxi-

mum value (999µm).

- Release the "+/-" button.

- Audible warning signal.

(When operating the instrument

for the first time or after the

E-EPROM has been exchanged.)

- The warning signal will cease

automatically after the initializa-

tion phase.