Setup the instrument – Leica Biosystems VT1000 A User Manual

Page 19

Advertising
background image

19

Leica VT1000 A

4. Setup the Instrument

Fig. 17

4.7 Assembling the specimen vice assembly

• The specimen vice assembly (34,

Fig. 16

) is

assembled inside the main specimen bath

(23,

Fig. 15

).

• Insert the cheese head screw (35) into the

bore of the specimen stage (27) and tighten

it using a

size 5

Allen key (37, included in the

scope of delivery).

• When tightening, ensure that the fixture (34)

is parallel to the front edge of the Vibratome

tray (

Fig. 17

).

• Insert the specimen block (36) into the ten-

sion clamp and clamp it into place using the

knurled screw (39) (

Fig. 18)

.

• The specimen to be sectioned is affixed to

the specimen block using the cyanoacrylate

adhesive provided.

• The anterior knurled screw, allows for orien-

tation of the specimen vice assembly, +/- 8

degrees. To change orientation, loosen the

screw and press on either side of the ten-

sion clamp. (refer also to

Fig. 22

).

Fig. 16

23

27

34

Fig. 15

35

36

37

37

34

Fig. 18

36

38

39

Always remove the blade BEFORE in-

stalling or removing accessories!