Leica VT1000 A

4. Setup the Instrument

43

Fig. 25

4.7.2 Assembling the rotating stage assembly

The rotating stage assembly allows for simple

position correction of the specimen surface

when the specimen is glued in place.
• To do so, first remove the specimen vice as-

sembly (34,

Fig. 16

) by following the reverse

order of steps as described in

Chap. 4.7

.

• Unscrew the knurled screw (38) completely.

You can now take the specimen clamp (40)

off of the base plate (39) (

Fig. 23

).

• Attach the rotating stage assembly (42) to

the base plate so that the pin (41) fits into the

bore (45) provided.

• Then, screw the knurled screw (38) into the

thread and tighten it, thus fastening the ro-

tating stage assembly (41) to the base plate

(39).

• Install it in the main specimen bath (23) as

described in

Chap. 4.7

(Fig. 25).

Fig. 23

38

39

40

39

42

41

45

Fig. 24

38

44

• The specimen to be sectioned (44) is glued

to the round specimen plate using the cy-

anoacrylate adhesive provided.

• To orient the specimen, slightly loosen the

knurled screw (43) on the right side, adjust to

the desired position and retighten the screw.