The operating elements

(continued)

THICKNESS switch

The THICKNESS switch defines the desired section thickness for au-

tomatic mode in microns.
• The parameter selected for this value is indicated in the LED dis-

play above the switch, the displayed values are specified in µm.

The LED (53) in the switch is then illuminated.

Switch position:
Upwards --> Values for the section thickness are increased.
Downwards --> Values for the section thickness are decreased.

54

SECTION switch

Fig. 35

Fig. 36

Fig. 37

Using the SECTION switch, the number of sections required for auto-

matic mode (AUTO) are defined.
• The value selected for this parameter is indicated in the LED display

above the switch; the LED (54) in the switch is then illuminated.

Switch position:
Upwards --> Number of sections is increased.
Downwards --> Number of sections is decreased.
• If this parameter is set to the value ‘000‘ manually, the instrument

switches to single sectioning mode until automatic mode is deacti-

vated via the switch.

Additional control operations available in automatic mode
• After the automatic sectioning process starts, after each section,

the LED display shows the number of sections remaining, based on

the value defined for the number of sections.

• If both switches, THICKNESS and SECTION, are pushed upwards

or downwards simultaneously, the specimen stage is moved up or

down to the maximum or minimum height position. When this posi-

tion is reached, the LED (54) in the THICKNESS switch flashes.