25

Leica VT1000 A

5. Operation

The operating elements

(continued)

Fig. 32

AMPLITUDE rotary knob

This knob controls the relative amplitude of the lateral excursion of

the sectioning blade vibratory movements.
• In the zero position (‘0‘) or during the retraction movement of the

cutting head, the blade does not vibrate.

• The maximum amplitude (2.0 mm) is reached when the rotary knob

is in the ‘10‘ position. At intermediate settings, the amplitude will

vary proportionally.

All configured values are active during the feed movement of

the cutting head only.

The amplitude can also be changed during the sectioning pro-

cess.

STAGE switch

Using this switch, you can move the specimen stage with the specimen

upwards or downwards quickly.
The movement continues as long as the switch is held in the respective

position. When the switch is released (mode switch must be in auto or

single position), it springs back into the center position and the move-

ment is interrupted.
Switch position:
Upwards --> Specimen stage moves upwards.
Downwards --> Specimen stage moves downwards.

Fig. 33

The LED display shows the values defined for section thickness in

single mode and the number of sections and section thickness in au-

tomatic mode.
• The LED (53, 54) in the THICKNESS (section thickness) switch or in

the SECTION (number of sections) switch is illuminated when the

corresponding parameter is being displayed (see

Fig. 37

).

LED display