Leica Biosystems Leica HER2 FISH System - 30 Test Instructions for Use TA9217 EN-CE-Rev_D 08/04/2013

English

2x2 Concordance Results Leica BOND-MAX System

Data was grouped as negative (<2.00) or positive (≥2.00) for a 2x2 analysis The observed

agreement for 300 samples between the two tests in a 2x2 analysis show a concordance of

99.33% (298/300) with a 95% CI of 97.61–99.92% for the Leica BOND-MAX System
The percentage Positive Agreement (sensitivity) or the ability of the Leica HER2 FISH System

- 30 Test to correctly identify Abbott Molecular PathVysion HER-2 DNA Probe assay positive

cases (the percentage of specimens scored positive by both the Leica HER2 FISH System -

30 Test and manual Abbott Molecular PathVysion HER-2 DNA Probe Kit out of all the Abbott

Molecular PathVysion HER-2 DNA Probe Kit positive cases) was 99.03% (102/103).
The percentage Negative Agreement (specificity) or the ability of the test to correctly identify

Abbott Molecular PathVysion HER-2 DNA Probe Kit negative cases (the percentage of

specimens scored negative by the Leica HER2 FISH System - 30 Test and Abbott Molecular

PathVysion

HER-2 DNA Probe Kit out of all the Abbott Molecular PathVysion HER-2 DNA Probe

Kit negative cases) was 99.49% (196/197). See Table 4.

Abbott Molecular PathVysion HER-2 DNA Probe Kit

Negative (<2.0)

Positive (≥2.0)

Totals

Leica HER2 FISH
System - 30 Test
Leica BOND-MAX

Negative (<2.0)

196

1

197

Positive (≥2.0)

1

102

103

Totals

197

103

300

Overall Concordance (95% CI) = 99.33% (97.61 – 99.92%)

Table 4. 2x2 concordance of Leica HER2 FISH System - 30 Test on the Leica BOND-MAX System with Abbott Molecular
PathVysion HER-2 DNA Probe Kit.