L. lot-to-lot precision study, Assay robustness – Leica Biosystems HER2 FISH System - 30 Test User Manual

Page 19

Advertising
background image

Page 19 of 24

English

Leica Biosystems Leica HER2 FISH System - 30 Test Instructions for Use TA9217 EN-CE-Rev_D 08/04/2013

English

K. Between Observer Precision Study

The between observer precision study was performed in a randomized and blinded fashion.

Between observer reproducibility testing of the Leica HER2 FISH System - 30 Test was

evaluated between three investigational sites. A single experienced observer at each

investigational site was used. Eighteen whole section breast cancer cases were used for

between observer precision, reflecting samples types used in the clinical setting.
On enumeration of the slides stained in the Between Observer Precision Study, 53/54 cases

evaluated demonstrated a concordant result giving an overall concordance of 98.15% with

a lower 95% CI of 90.11%.

L. Lot-to-Lot Precision Study

The lot-to-lot precision study was performed in a randomized and blinded fashion. Lot-to-Lot

precision was determined on three independently manufactured lots of Leica HER2 FISH

System - 30 Test, manufactured under Good Manufacturing Practice (GMP). Each lot was

tested at a single investigational site on 540 previously HER2 characterized TMA samples

containing formalin-fixed paraffin-embedded breast cancer cases. The use of TMAs for the

determination of lot-to-lot reproducibility enables a larger volume of cases covering a wider

range of HER2 expression to be tested between lots.
On enumeration of the slides stained in the Lot-to-Lot Precision Study, 540/540 cases

evaluated demonstrated a concordant result giving an overall concordance of 100% with a

lower 95% CI of 99.45%.

Assay Robustness

Robustness studies were performed on the Leica BOND-MAX and BOND-III System to determine

the assay tolerance range for heat retrieval time and temperature; enzyme retrieval

time, temperature and concentration; denaturation time and temperature; hybridization time and

temperature; and stringency wash time and temperature. Robustness studies using the default

Leica BOND-MAX and BOND-III System protocol were also performed outside the

recommended limits as defined in the FDA/ORA guidance document ORA LAB5.3 Rev1.7 for

temperature and humidity.

• No difference in amplification status was observed when the default temperature for each

heat dependent step was raised by 4 °C or decreased by 4 °C, when compared to the default

Leica HER2 FISH System - 30 Test protocol. Highest quality ratings were observed at the default

temperatures and these temperatures are recommended.

• No difference in amplification status was observed when the heat induced epitope retrieval

(HIER) time was performed for 20 minutes and 30 minutes at 97 °C with Leica BOND ER1

solution, when compared to the default Leica HER2 FISH System - 30 Test protocol. Highest

quality ratings were observed at the default time of 25 minutes and this incubation time is

recommended.

• No difference in amplification status was observed when the enzyme induced epitope retrieval

(EIER) time was performed for 15 minutes and 35 minutes at 37 °C, when compared to the

default Leica HER2 FISH System - 30 Test protocol. Highest quality ratings were observed at

the default time of 25 minutes and this incubation time is recommended.

• No difference in amplification status was observed when the enzyme induced epitope retrieval

(EIER) enzyme concentration was performed with enzyme concentrate/enzyme diluent

ratios of 1:200 and 1:500 using the default Leica HER2 FISH System - 30 Test protocol.

Highest quality ratings were observed at the default concentration of 1:300 and this dilution

is recommended.

• No difference in amplification status was observed when the denaturation time was performed

for 5 minutes and 15 minutes, when compared to the default Leica HER2 FISH System - 30

Test protocol. Highest quality ratings were observed at the default time of 10 minutes and this