Leica Biosystems Leica HER2 FISH System - 30 Test Instructions for Use TA9217 EN-CE-Rev_D 08/04/2013

F. Lot-to-Lot Precision Study

The lot-to-lot precision study was performed in a randomized and blinded fashion. Lot-to-Lot

precision was determined on three independently manufactured lots of the Leica HER2 FISH

System - 30 Test, manufactured under Good Manufacturing Practice (GMP). Each lot was

tested at a single investigational site on 540 previously HER2 characterized TMA samples

containing formalin-fixed paraffin-embedded breast cancer cases. The use of TMAs for the

determination of lot-to-lot reproducibility enables a larger volume of cases covering a wider

range of HER2 expression to be tested between lots.
On enumeration of the slides stained in the Lot-to-Lot Precision Study, 534/540 cases

evaluated demonstrated a concordant result giving an overall concordance of 98.89% with

a lower 95% CI of 97.60%.

Precision Testing – Leica BOND-III System

G. Within Run Precision Study

The within run precision study was performed in a randomized and blinded fashion. Within

run precision testing of the Leica HER2 FISH System - 30 Test was evaluated at a single

investigational site on 540 previously HER2 characterized TMA samples containing formalin-

fixed paraffin-embedded breast cancer cases. The use of TMAs for the determination of

within run precision enabled a larger volume of cases covering a wider range of HER2

expression within a single run on a single instrument.
On enumeration of the slides stained in the Within Run Precision Study, 540/540 cases

evaluated demonstrated a concordant result giving an overall concordance of 100% with a

lower 95% CI of 99.45%.

H. Within Instrument Precision Study

The within instrument precision study was performed in a randomized and blinded fashion.

Within instrument precision testing of the Leica HER2 FISH System - 30 Test was evaluated

at a single investigational site on 1620 previously HER2 characterized TMA samples

containing formalin-fixed paraffin-embedded breast cancer cases. The use of TMAs for the

determination of within instrument precision enabled a larger volume of cases covering a

wider range of HER2 expression within multiple runs on a single instrument.
On enumeration of the slides stained in the Within Run Precision Study, 1620/1620 cases

evaluated demonstrated a concordant result giving an overall concordance of 100% with a

lower 95% CI of 99.82%.

I. Between Run Precision Study

The between run precision study was performed in a randomized and blinded fashion.

Between run precision testing of the Leica HER2 FISH System - 30 Test was evaluated

at a single investigational site on 900 previously HER2 characterized TMA samples

containing formalin-fixed paraffin-embedded breast cancer cases. The use of TMAs for the

determination of between run, day-to day precision testing enabled a larger volume of cases

covering a wider range of HER2 expression to be tested between runs on different days.
On enumeration of the slides stained in the Between Run Precision Study, 891/900 cases

evaluated demonstrated a concordant result giving an overall concordance of 99.00% with

a lower 95% CI of 98.11%.

J. Between Laboratory Precision Study

The between laboratory precision study was performed in a randomized and blinded

fashion. Between laboratory precision testing of the Leica HER2 FISH System - 30 Test

was evaluated between three investigational sites on 513 previously HER2 characterized

TMA samples containing formalin-fixed paraffin-embedded breast cancer cases. The use of

TMAs for the determination of between laboratory precision testing enabled a larger volume

of cases covering a wider range of HER2 expression to be tested between runs on multiple

instruments.
On enumeration of the slides stained in the Between Laboratory Precision Study, 511/513

cases evaluated demonstrated a concordant result giving an overall concordance of 99.61%

with a lower 95% CI of 98.60%.