Troubleshooting

Problem

Probable Cause

Remedial Action

No or weak
fluorescent
signal/staining

Inappropriate fixation or
processing of test specimen

Ensure a formalin-based fixative is used and
that processing schedules are suitable for
the specimen undergoing testing.

Leica HER2 FISH System - 30
Test is being used outside its
expiry date

Ensure the Leica HER2 FISH System - 30
Test used is within its specified expiry date.

Incorrect protocol selection

Ensure appropriate default to *FISH Protocol
A in the staining protocol field of the Add slide
dialog.

Inappropriate bulk reagents
dispensed

Ensure all Leica BOND reagents have been
allocated to appropriate bulk containers
and placed into appropriate positions on the
instrument.

Inadequate deparaffinization
of slides

Ensure *Dewax mode is selected in the
Preparation field of the Add slide dialog.

Inappropriate pretreatment

Ensure default pretreatment (HIER and
Enzymatic Digestion) protocols are selected.
Adjust pretreatment protocol (HIER or
Enzymatic Digestion) if required.

Inadequate denaturation

Ensure appropriate default denaturation *D10 is
selected.

Inadequate hybridization

Ensure appropriate default hybridization *H12 is
selected. Extend hybridization time if required.

Excess post hybridization
washing

Decrease post hybridization wash incubation
time.

Run aborted prior to
completion

Using Leica BOND software, confirm the
presence of any reportable errors during the
staining run and address as instructed by
the Leica BOND software.

Inappropriate fluorescence
microscopy equipment

Inappropriate filter set

Incorrect lamp

Expired lamp

Incorrect oil type

Ensure that all fluorescence microscopy
equipment used is appropriate for the assay
being performed, confirm:

Appropriate filter set

Appropriate lamp

Good lamp strength

Appropriate oil for use in oil immersion
microscopy

Over exposure to UV light
(photobleaching)

Store slides before and after assessment in
the dark to preserve fluorescent signals. To
maintain signal for long term storage store
slides at –20 ºC.