Instrument

Magnification

stage

Lens

Tube factor

Eyepieces

Graticule

Calibration value

Table

Calibrating
1.
Focus stage micrometer.

2. Count:

number of mm (inches) on the stage micrometer (X)

number of intervals on the eyepiece graticule (Y)

3. Calculate calibration value (one interval of the eyepiece graticule):

X

Y

= mm (inch) calibration value

Measuring
4.
Focus on specimen rather than stage micrometer.

5. Count and calculate:

number of intervals on the eyepiece graticule

mm (inch) calibration value = mm (inch) length of measurement distance

×