1. Measuring pitch

12mm : 120

Stage micrometer

with 0.1mm

graduation

Examples

Calibrating

7.8mm on the stage micrometer (X)

120 intervals on the eyepiece graticule (Y)

7.8

120

mm = 0.065mm calibration value

Measuring

Intervals on the eyepiece graticule: 112

Result:

112 intervals

× 0.065mm = 7.3mm length of

measurement