Leica Biosystems M80 Measuring User Manual

Page 7

Advertising
background image

7

Measuring – User Manual

̈ Place stage micrometer on

the table insert and set
sharpness.

̈ Select the magnification

with which you will later
want to take measurements.

̈ Set eyepiece graticule and

lens micrometer to be free of
parallax (see page 4).

̈ Align stage micrometer (X) close to and in parallel

with the eyepiece graticule (Y).

̈ Focus both pitches down to a single point.

In the example: 0 and 10

̈ Count: The number of mm (inches) of the stage

micrometer (X) corresponds to a certain number of
intervals on the eyepiece graticule (Y).

̈ In the example:

7,8mm on the stage micrometer corresponds to 120
intervals on the eyepiece graticule

̈ Calculate calibration value using calibration formula

and note down.

X stage micrometer: number of mm (inches)

Y eyepiece graticule: number of intervals

Calibrating

Calibration formula