Measuring – User Manual

̈ Select the lowest

magnification.

̈ Observe specimen through the

eyepiece with graticule and
focus.

̈ Select the highest

magnification.

̈ Use focusing drive to

optimise sharpness.

Do not look into the eyepieces
̈ Turn eyepiece without grid

plate anticlockwise until
it reaches the stop.

̈ Select the lowest

magnification.

̈ Observe specimen through the

eyepiece without graticule and
slowly turn eyepiece clockwise
until the eye clearly sees the
specimen.

̈ Select the highest

magnification.

̈ Observe specimen through

both eyepieces.

̈ If necessary, use the focusing

drive to re-focus.

̈ Look into the eyepieces and slightly move your head.
• Graticule and specimen must not move towards one

another, i.e. they must be clearly focused in one plane
(free of parallax). If this is not the case, please repeat
the process.

Regulate

eyepiece without

graticule

Checking eyepiece

is free of parallax