Transmitted-light Base TL BFDF

Standard delivery

The base is delivered with the adapter plate
installed. The selected stage (IsoPro™ cross-
stage or standard stage 10 447 269), and the
focusing drive will have to be mounted later.

Stage assembly

The TL BFDF transmitted-light base can be
equipped with three different stages.

Leica IsoPro™ manual cross-stage

Leica IsoPro™ automatic cross-stage

Standard stage 10 447 269

The selected stage is mounted on the base
before commissioning. You can switch between
the stages at any time with just a few hand
movements.

The following paragraph assumes use of the
base without the stage mounted. Disassembly
is performed in reverse order of the following
steps.

Standard stage

1. Take the glass plate from the rectangular

gap in the standard stage.

2. Position the stage on the transmitted-light

base in such way that the four holes align
over those in the base.

3. Attach the stage to the base using the four

Allen screws provided.

4. Insert the glass plate back into the standard

stage.