Objective nosepiece – adjusting parfocality – Leica Biosystems M205A User Manual

Page 26

Advertising
background image

Leica M series

Manual

26

Objective Nosepiece – Adjusting Parfocality

The following procedure only has to be carried
out once. Afterwards, both objectives are parfo-
cal, meaning that the specimen remains in focus
when the objective nosepiece is rotated.

This procedure must be repeated if you replace
either of the two objectives with another.

The following example assumes the combi-
nation of the 1× and 2× planapochromats. If
you are using another objective combination,
replace the 2× objective in the description with
the objective with the stronger magnification.

Preparation

Open the iris diaphragm.

Set the dioptric correction of the eyepieces

to "0".

Adjustment

1. Rotate the 2× objective into the beam path

and set it to the lowest magnification.

2. Focus on the specimen.

3. Rotate the 1× objective into the beam

path.

4. Turn the objective on the thread in both

directions until the specimen appears
sharp.

5. Toggle to the 2× objective.

6. Select the strongest magnification and

refocus until the specimen appears abso-
lutely sharp.

7. Toggle to the 1× objective.

8. Turn the objective on the thread in both

directions until the specimen appears
absolutely sharp.

By means of zooming, check that the behavior
of the objective is parfocal. Repeat the check
with the other objective. If it is not parfocal,
repeat the procedure.

9. Tighten the locking screws.

Advertising
This manual is related to the following products: