Safety regulations (continued)

Direct contact with eyepieces can be a potential
transmission path for bacterial and viral infec-
tions of the eye.

The risk can be kept to a minimum by using
personal eyepieces for each individual or
detachable eyecups.

Light sources: safety instructions

Protective measures of the manufacturer:

UV protection screen in front of the speci-

★

men plane prevents the user from looking
directly into the UV rays.

Dummy filter carriers in the free positions

★

of the rapid filter changer prevent direct UV
radiation from reaching the eyes.

UV filters are installed in the observation

★

beam paths to protect the eyes.

The stray-light protection on the lamp

★

housing prevents irradiation of the hands.

Warning

UV radiation could damage the eyes. Therefore:

Never look into the light spot on the speci-

★

men plane without a UV protection screen.

Never look into the eyepieces if no excita-

★

tion filter is in the beam path.

Fill empty filter positions with dummy filter

★

carriers (e.g. M165FC).

Do not select a white, strongly reflective

★

background for the stage.