IsoPro™ Manual Cross-stage: Assembly

Before the IsoPro™ cross-stage is mounted to
the base, the axis containing the control buttons
is attached either on the left or the right side of
the cross-stage.

IsoPro™ Cross-stage

Before the IsoPro™ cross-stage is mounted to
the base, the axis containing the control buttons
is attached either on the left or the right side of
the cross-stage.

If the controls are to be mounted on the left-
hand side, the gear rod on the bottom side of
the cross-stage must be unscrewed and reat-
tached in reverse.

1. Take the glass plate from the cross-stage.

2. Turn the cross-stage around and place it

onto a non-slip surface.

3. Change the gear rod from the left to the

right-hand side.

4. Skip the next two steps to mount the

controls.

Left or right operation

If the controls are to be mounted on the left-
hand side, the gear rod on the bottom side of
the cross-stage must be unscrewed and reat-
tached in reverse.