Leica SmartTouch™

Manual

24

Registering components

1. Touch the "Config." tab.

2. In the left area, touch the symbol for the

microscope accessories.

3. Touch the symbol of the instrument compo-

nent that you would like to detect.

 

Unencoded components (such as an
objective) are not detected automati-

cally by the Leica SmartTouch™. To ensure that
these components are included in the various
calculations (depth of field, zoom factor, etc.)
correctly, they must be configured in the Leica
SmartTouch™.

 

These configurations must be carried out
on the following occasions:

Ϙ

Initial commissioning of the microscope.

Ϙ

After an unencoded component has been
replaced by a component with different
properties.

Ϙ

After the Leica SmartTouch™ has been reset
to the factory defaults.

4. In the list with the available components,

touch the components you are using.

5. Repeat steps 3 and 4 for all components

that you would like to use.

 

Each change is stored immediately.
Restarting the instrument is not required.

Configuration of Unencoded Components