background image

DQ-4000

DQ-4000 User's Guide

Page: 19 / 21

P/N : 033-02061 Rev. 1

4.9 SLC (Load Capacity) Command

Function:

Sets (or reports) load capacity of the load cell on the selected channel

Summary:

Sets or reports the load capacity of the currently selected channel using S1, S2,
S3 or S4 command. If S0 command was previously issued, SLC would return
the total load capacity of all channels.

Example:

SLC 50 (sets load capacity to 50 lb)

SLC (reports current load capacity in lb)

4.10 CN0, CN1, CP0, CP1, CAZ, FAZ, CLA & CTA Commands

Function:

Functions for manufacturer use only—

do not change!

Summary:

These commands are used to set parameters that affect the calibration of the
sensors and are not to be changed except under the guidance of Loadstar
Technical Support. The DQ-4000 will report erroneous load, if these
parameters are not set correctly.

4.11 CK0 Command

Function:

Sets (or reports) the current compensation factor

Summary:

If you purchased your iLoad Mini (or other Loadstar frequency-output load cell)
separately from the DQ-4000, you may be provided with a value of K0. Use this
command to enter the value of K0 for the sensor connected to the selected
channel.

Example:

CK0 1.43576 (sets K0 to be 1.43576)