MTS Tension Compression Grips User Manual

Page 9

Advertising
background image

Operation

2-3

10.

Slide the upper collar over the end of the specimen as shown
in Figure 2-2 and let it rest on the lower collar.

11.

Install the split button-end adapter into the upper collar as
shown in Figure 2-2.

WARNING

Failure to zero dc error will result in sudden actuator
movement.

This can cause injury to personnel and/or damage to
e q u i p m e n t .

Do not apply system hydraulic pressure with other than zero dc error
(command = feedback).

12.

Apply full system hydraulic pressure according to applicable
system procedures. Use extreme care not to cause rapid
actuator movement when performing the following steps.

Place upper

collar and

adapter over

specimen and

rest them on

lower collar

Extend actuator until

specimen contacts

upper base

Ensure equal

measurement all

the way around

Slide collar and
split button-end

adapter up

specimen to

upper base

Ensure equal

measurement all

the way around

Secure with socket

head cap screws