Surge suppressor kit, Surge suppressor kit 65 – MTS Series 505G2 SilentFlo Hydraulic Power Unit Model 505G2-11 User Manual

Page 65

Advertising
background image

Surge Suppressor Kit

Model 505G2.07/.11 SilentFlo™ HPU

65

5. Screw the filter bowl back onto the filter assembly. Torque the filter bowl to

27.1 N•m (20 lbf-ft).

6. Press the Reset button on the HPU control panel to reset any interlocks and

turn off the Dirty Filter indicator.

Surge Suppressor Kit

The surge suppressor kit smooths out the ripple caused by the pulsing action of
the pump as it pressurizes the hydraulic fluid. The surge suppressor is a cylinder
that has an outer sleeve that is pressurized with nitrogen. The gas pressurized
sleeve dampens the pressure pulses.

The surge suppressor is mounted to the output of the hydraulic pump inside the
reservoir.

Required equipment

You will need the following to install the Surge Supressor kit:

A Model 505.07 Surge Suppressor kit (part number 100-008-667) or a
Model 505.11 Surge Suppressor kit (part number 100-008-665).

Optional Remote Charging kit (part number 100-056-912)

A transfer pump with a 10-micron filter (such as the Model 590.01 Fluid
Transfer Pump).

A cleaning solvent such as ethanol.

A charging kit (part number 037-698-601).

Installation procedure

To install the Surge Suppressor Kit.

Remote Charging adapter