MTS 318 Load Unit User Manual

Page 85

Advertising
background image

Routine Maintenance Overview Checklist

318 Load Unit

Maintenance

85

4. Use the Lock Control to lock the crosshead. Then torque the manual

crosshead locking bolts in the order shown in the following illustration.

5. Turn off hydraulic pressure.

6. Use the Lock Control to the unlock crosshead position to remove pressure

from the hydraulic locks. Wait two minutes for the pressure in the locks to
drop to zero before going on to the next step.

7. Tighten each lock’s cap screw until its piston bottoms out. Then loosen and

hand-tighten each cap screw.

8. Loosen each of the hydraulic locks’ cap screws 1/4 turn.

9. Turn on electrical power at the test controller if you have not already done

so.

10. Reset any active interlocks at the test controller.

11. Turn on high hydraulic pressure.

12. Return the Lock Control to the lock crosshead position to pressurize the

hydraulic locks.

13. Fully loosen the manual crosshead locking bolts. Then turn the Lock

Control to the unlock crosshead position.

Tighten

Bottom out
position

Loosen, then
hand tighten

Loosen1/4 turn

Advertising