Set up the test, Fatigue crack growth (fcg) templates overview, Set up a test – MTS Fracture Template User Manual

Page 26

Advertising
background image

Set Up the Test

Fatigue Crack Growth (FCG) Templates Overview

There are two versions of Fatigue Crack Growth templates. Both templates comply with ASTM standard
E647-08.

ASTM E 647 Fatigue Crack Growth Compliance

ASTM E 647 Fatigue Crack Growth DCPD

The primary difference between the two templates is how the crack size is measured. The compliance
template uses a Crack Opening Displacement (COD) clip gage to measure the crack opening. The DC
Potential Drop (DCPD) template uses electrical leads that carry voltage, and the change in voltage relates
to a change in crack size.

Control modes

The template guides you through the steps of running a test and provides all the software components
you need to run a fatigue crack growth test in one of two control modes:

Constant Load—The test cycles between a specified minimum and maximum load.

Delta K—The test cycles between the minimum and maximum loads that correspond to the desired
minimum and maximum stress intensities (K).

Features

The main features of the FCG template are listed below:

The main template window guides you through setting up all test parameters, performing crack size
checks, precracking the specimen, and running the test.

As the test runs, you can stop the test to change test parameters.

Comprehensive monitor views help you monitor test progress.

Tabular and graphical displays make it easy to review test results.

Analysis definitions provide all necessary analysis calculations and views to analyze test runs.

Report templates help you generate reports for analysis runs.

Set Up a Test

To create and run a new test:

1. Create a new test from a template.

a) Click File > New > Test from Template.
b) In the Create from Existing Test window, select a test.
c) Click OK.

26 | MTS TestSuite

Fatigue Crack Growth (FCG) Template

Advertising