Assign precrack values, Calculator – MTS Fracture Template User Manual

Page 80

Advertising
background image

Assign Precrack Values

Specify these parameters if the specimen was precracked on another load frame. When the precrack
parameters are specified here, you can then skip Precrack Specimen in the Main Menu window.

Precrack Parameters

Description

Parameter

Specify the number of precrack cycles
completed.

Precrack Cycles
Completed

Specify the final crack size calculated from
compliance.

Precrack Final Crack Size

Specify the final load during precrack.

Precrack Final P

Specify the maximum load measured during
precrack.

Precrack P Maximum

Specify the final K during precrack.

Precrack Final K

Specify the maximum K value during
precrack.

Precrack K Maximum

Specify any additional comments.

Precrack Comments

Calculator

Use the Calculator to calculate the load to use for new specimens. You can also use the Calculator to
calculate stress intensity (K). Load is calculated from the crack size value and stress intensity (K) value.
Stress intensity (K) is calculated from the crack size value and load value.

Typically, you use the Calculator after the crack size is checked. If the crack size check is performed, the
Multipurpose Elite application supplies the crack size value for the Calculator. You can also enter a crack
size value manually. You must enter the load or stress intensity value manually.

Calculate Load P

Load P Parameters

Description

Parameter

Specify the crack size value to use in the load
calculation.

Crack Size

80 | MTS TestSuite

Run the CTOD Fracture Toughness Test

Advertising