a) Click New Test Run button.
b) Select a specimen from the Select a Specimen window and then click OK.

To create a new specimen click Add a new item.

c) Review the variables in the Setup Variables window, modify values as necessary, and then click

OK.

3. Apply hydraulic power to the system.

a) Reset the interlocks if needed.
b) Click the Low, then High, power buttons.

4. In the MTS TestSuite Custom Message window for the template, click the buttons to define test

parameters, perform setup tasks, and run tests.

5. When prompted, click Run to run the test.

Define Test Parameters

Test Parameters

After a new test run is added to the test, the Main Menu window appears. This window provides access
to all necessary test parameters. Click the parameter buttons in the Main Menu window to specify parameter
values for the test.

Precrack Parameters

Precrack Parameters

Description

Parameter

Specify the desired crack size to achieve in
the precrack activity. When the specified
crack size is achieved, the activity stops.

Precrack Final Crack
Limit

Specify the cycle frequency of the command
signal for the precrack activity.

Precrack Frequency

Specify the ratio of the minimum to maximum
load applied to the specimen. The minimum

Precrack Load Ratio

load is determined by this value and the
maximum load is specified by you.

Specify the percentage of the measured load
range used as the lower limit for linear least

Precrack Lower Least
Squares Fit Percentage

squares fit compliance determination. This
parameter is not used for FCG DCPD tests.

MTS TestSuite | 63

KIC Fracture Toughness Template

Advertising