Ctod parameters, Data storage parameters properties – MTS Fracture Template User Manual

Page 78

Advertising
background image

Description

Parameter

Specify the percentage of the measured load
range used as the upper limit for linear least

Precrack Upper Least
Squares Fit Percentage

squares fit compliance determination. This
parameter is not used for FCG DCPD tests.

Specify the percent of maximum load level
at which the voltage is measured. The load

Precrack Measure Load
Level Percent

is applied at this level to ensure the crack is
open. All voltage data below this load will not
be used to calculate crack length. This
parameter is used only for FCG DCPD tests.

Specify the number of cycles at which
precrack activity stops.

Precrack Cycle Limit

Specify the maximum stress intensity desired
at the end of precrack activity. The initial

Precrack Final Maximum
K

precrack Maximum K is 1.4 times the Final
Maximum K value.

Select Yes to cause an interlock at the end
of precrack. This can shut down hydraulics,
if desired.

Shutdown Station at End
of Precrack

CTOD Parameters

In the CTOD Parameters section of the Main menu, you select COD or Displacement as the control
mode
for the test, and then specify the Ramp Rate, which is the rate for the ramp segment of the CTOD
step.

Data Storage Parameters Properties

Data Storage Parameters

Description

Parameter

Specify the percentage of precrack activity
that must complete before data is saved to
disk.

Precrack Save Percent
Limit

Specify the cycle intervals written to disk
during the precrack activity.

Store Every Nth Precrack
Cycle

78 | MTS TestSuite

Run the CTOD Fracture Toughness Test

Advertising